Two Indenters Vickers Hardness Tester

Two Indenters Vickers Hardness Tester

Details
QHVT-10ZD/50ZD
Built with a new aluminum alloy die-casting process, the QHVT-10ZD/50ZD digital Vickers and Knoop hardness tester boasts a sturdy, stable housing with a sleek, modern design. Its direct drive loading mechanism ensures consistent force application and high testing throughput. The unit comes with two indenters and four objectives that feature automatic recognition and switching, making it adaptable to a variety of testing scenarios. A digital eyepiece works in conjunction with the touch screen to clearly display test force, indentation length, dwell time, conversion scale, date, and time. Operation is straightforward: after measuring the indentation, simply press the eyepiece button, and the hardness value is automatically calculated and shown on the screen.
Category
Vickers Hardness Tester
Share to
Send Inquiry
Description
Technical Parameters

Introduction

Built with a new aluminum alloy die-casting process, the QHVT-10ZD/50ZD digital Vickers and Knoop hardness tester boasts a sturdy, stable housing with a sleek, modern design. Its direct drive loading mechanism ensures consistent force application and high testing throughput. The unit comes with two indenters and four objectives that feature automatic recognition and switching, making it adaptable to a variety of testing scenarios.

A digital eyepiece works in conjunction with the touch screen to clearly display test force, indentation length, dwell time, conversion scale, date, and time. Operation is straightforward: after measuring the indentation, simply press the eyepiece button, and the hardness value is automatically calculated and shown on the screen.

product-800-450
  • Crafted with a new aluminum alloy die-casting process, the main body is sturdy, stable, and visually appealing.
  • Crafted from a specialized material and heat-treated to 35–40HRC, the lifting screw offers exceptional wear resistance. The worm gear and worm structure further ensures smooth, stable lifting operation.
  • The closed-loop load cell loading system, combined with a direct-drive motor that applies test force vertically, provides exceptionally stable performance.
  • The turret's rapid rotation significantly enhances overall testing productivity.
  • The loading and turret motors are both imported, ensuring superior build quality, low noise levels, and reliable stability.
  • The modular cover design of the machine body ensures hassle-free maintenance.
  • With a six-station turret that holds Vickers and Knoop dual indenters and 5X/10X/20X/40X objectives, the instrument automatically identifies and switches between them for seamless operation.
  • The turret switching system integrates mechanical and photoelectric limits, providing superior positioning accuracy to maintain precise six-point-one-center alignment.
  • The LED light source delivers superior durability for prolonged use.
  • The 8-inch HD color touch screen features an adjustable angle, making operation more comfortable and user-friendly.
  • With Vickers and Knoop testing modes built in, the system allows for real-time conversion between various hardness scales and convenient on-demand switching.
  • The system allows for upper and lower limit configuration, delivering clear prompts whenever test results exceed the specified tolerance.
  • The instrument offers real-time storage of hardness test results, with functionality for data classification and selective deletion. Automatic calculation of maximum, minimum, and average values is provided. Furthermore, an RS232 interface is included to facilitate PC connectivity for data output.

 

Application Range

 

This equipment is suitable for a broad spectrum of materials: ferrous and non-ferrous metals, IC thin sections, coatings, ply-metals, glass, ceramics, agate, precious stones, and thin plastic sections. It is also applicable for hardness testing of carbonized layers and quench-hardened layers, with capabilities to evaluate both depth and trapezoidal profiles.

 

Detail Pictures

product-488-366
Operation interface
product-642-473
Two indenters, Four objectives

Technical Parameters

 

Model

QHVT-10ZD

QHVT-50ZD

Test Force

kgf

0.3kgf, 0.5kgf, 1kgf, 2kgf, 2.5kgf, 3kgf, 5kgf, 10kgf

0.5kgf, 1kgf, 2kgf, 2.5kgf, 3kgf, 5kgf, 10kgf, 20kgf, 30kgf, 50kgf

N

2.94N, 4.90N, 9.80N, 19.6N, 24.5N, 29.4N, 49N, 98N

4.90N, 9.80N, 19.6N, 24.5N, 29.4N, 49N, 98N, 196N, 294N, 490N

Test Mode

HV/HK

Test Range

1HV~3000HV

Hardness Resolution

0.1HV

Loading Method

Automatic (Loading/Dwell/Unloading)

Turret

Automatic Shifting

Data Output

RS232 Interface

Hardness Reading

8 inch Color Touch Screen, Automatic Data Storage

Conversion Scale

HRA, HRBW, HRC, HRD, HREW, HRFW, HRGW, HRHW, HRKW, HR15N, HR30N, HR45N, HR15TW, HR30TW, HR45TW, HV, HK, HS, HBS, HBW, HLD

Data Storage

6 folders, each 1000 sets data

Eyepiece

10× Digital Measuring Eyepiece

Objective

5×,10×,20×,40×

Total Magnification

50×,100×,200×,400×

Effective Field of View

5×:1.6mm,10×:800μm,20×:400μm,40×:200μm

Min. Measuring Unit

5×:0.5μm,10×:0.25μm,20×:0.13μm,40×:0.06μm

Dwell Time

0~99s

Light Source

LED

Digital X-Y Test Table

Size: 120×120mm; Travel: 25×25mm; Resolution: 0.001mm

Max. Sample Height

175mm

Throat Depth

140mm

Power Supply

AC220V,50Hz

Execute Standard

GB/T 4340,ISO 6507,ASTM E92,JIS Z2244

Dimension

495×575×670mm

Net Weight

35kg

 

Packing List

 

Name

Qty

Name

Qty

Instrument Main Body

1 set

Accessory Box

1 pc

10× Digital Measuring Eyepiece

1 pc

Digital X-Y Test Table

1 pc

Standard Vickers Hardness Block

2 pcs

Standard Knoop Hardness Block

1 pcs

Horizontal Regulating Screw

4 pcs

Level

1 pc

Inner Hexagon Spanner 2.5mm, 3mm

each 1 pc

Screw Driver

2 pcs

Fuse 1A

2 pcs

Power Cable

1 pc

Anti-dust Cover

1 pc

Usage Instruction Manual

1 copy

 

FAQ

Q: How should fixtures be selected and maintained?

A: The selection of fixtures depends on the shape, material, and testing type of the sample. We offer a variety of conventional fixtures, and can also customize matching fixtures according to the shape of the sample. During maintenance, it is necessary to keep the fixtures clean, avoid collisions, and regularly apply rust proof oil.

Q: What should be done if abnormal noise or vibration occurs during equipment operation?

A: Please press the emergency stop button immediately to stop the test. Firstly, check if the equipment is placed horizontally and if the sample is placed correctly. If the problem persists, please contact our after-sales service department promptly.

Q: How to perform daily maintenance and upkeep on equipment?

A: Routine maintenance includes: keeping the equipment clean; regularly inspecting transmission components and adding appropriate lubricant; avoiding overloading the equipment; and periodically powering it on during long-term inactivity. Please refer to the accompanying "User and Maintenance Manual" for specific maintenance cycles and methods.

 

Hot Tags: two indenters vickers hardness tester, China two indenters vickers hardness tester manufacturers, suppliers, factory

Send Inquiry
Contact usif have any question

You can either contact us via phone, email or online form below. Our specialist will contact you back shortly.

Contact now!